Automatic field map generation and off-resonance correction for projectionreconstruction imaging

Citation
Ks. Nayak et Dg. Nishimura, Automatic field map generation and off-resonance correction for projectionreconstruction imaging, MAGN RES M, 43(1), 2000, pp. 151-154
Citations number
14
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MAGNETIC RESONANCE IN MEDICINE
ISSN journal
07403194 → ACNP
Volume
43
Issue
1
Year of publication
2000
Pages
151 - 154
Database
ISI
SICI code
0740-3194(200001)43:1<151:AFMGAO>2.0.ZU;2-D
Abstract
A new projection reconstruction technique utilizes the oversampling of low spatial frequencies to estimate and correct for off-resonance effects. Inte rleaved spokes are acquired at one of two different echo times. From separa ted early-TE and late-TE raw data, two one-quarter resolution images are re constructed and a one-quarter resolution field map is computed, Multifreque ncy reconstruction with all the data is then used to simultaneously correct for off-resonance and compensate for the difference in echo times. Resulti ng images obtained on phantoms and in vivo demonstrate significantly reduce d off-resonance artifact without the acquisition of a separate field map. M agn Reson Med 43:151-154, 2000, (C) 2000 Wiley-Liss, Inc.