Microchannel plate resistance at cryogenic temperatures

Citation
P. Roth et Gw. Fraser, Microchannel plate resistance at cryogenic temperatures, NUCL INST A, 439(1), 2000, pp. 134-137
Citations number
13
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
439
Issue
1
Year of publication
2000
Pages
134 - 137
Database
ISI
SICI code
0168-9002(20000101)439:1<134:MPRACT>2.0.ZU;2-N
Abstract
The resistance R of a "low-resistance" microchannel plate (MCP) electron mu ltiplier was measured over the extended temperature range 4-293 K. The rati o R(T)/R(293 K) reached values of 10(2), 10(3) and 10(6) at temperatures T = 59, 40 and 19 K, respectively. These results are interpreted in terms of variable range hopping conduction mechanisms in the reduced lead silicate M CP glass. The results may be important for cryogenic charged particle detec tion using MCPs, as, for example, in a recent proposal for a quantum comput er. (C) 2000 Elsevier Science B.V. All rights reserved.