Tv. Glotzer et al., Variability of atrial sensing during exercise in a patient with a dual chamber ICD caused inappropriate ICD discharges, PACE, 22(12), 1999, pp. 1838-1841
Citations number
10
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
We present the case of a patient with a dual chamber implantable cardiovert
er defibrillator (ICD) who experienced inappropriate ICD discharges during
exercise. Interrogation of the ICD revealed intermittent atrial undersensin
g during exercise that was responsible for the erroneous classification by
the ICD of sinus tachycardia as ventricular tachycardia. Monitoring of the
intracardiac electrograms and Marker Channels during an exercise test confi
rmed a marked decrease in P wave amplitude during exercise. By increasing t
he atrial sensitivity setting the problem wets resolved.