The First Order Kernel responses (FOK) from a group of 15 normal subjects a
ged 21-36 yrs were recorded under clinical conditions with the multifocal E
RG (MERG) systems VERIS and RETiscan using system-specific software and har
dware under Various parameter settings. First, the multifocal ERG's of norm
al subjects were recorded with the standard-parameters of VERIS and RETIsca
n. Thereafter the VERIS system parameters were set as close as possible to
the RETIscan-setup, and the recordings were repeated. The VERIS parameters
changed were the luminance- and contrast-settings of the stimulus-monitor a
nd m-sequence-timing. In addition, the effects of different sampling rates
were also examined. The consequences of the parameter changes were analyzed
by displaying the peak times of the First-Order-Kernel (P1). The parameter
-combinations used for recording lead to system-specific results. By varyin
g the luminance/contrast settings and/or the m-sequence-timing the results
can be changed. The amplitude of the recorded results can be influenced by
the luminance-settings of the stimulus-monitor, and the topography of the P
1 peak times is affected by the m-if-frame-combination of the recording sti
mulus. With their standard parameters, the two systems give different measu
res of the PI peak times. By adjusting the recording parameters. these diff
erences can be reduced significantly. The main parameter here is the m-sequ
ence-timing, although the sampling rate must also be taken into account wit
h regard to the accuracy of the recording results.