J. Bunkenborg et al., Doubling the sensitivity of natural abundance C-13-C-13 INADEQUATE with off-resonance compensation, MAGN RES CH, 38(1), 2000, pp. 58-61
The INADEQUATE CR experiment about doubles the sensitivity of the original
INADEQUATE experiment but it still suffers from severely deteriorated perfo
rmance in the presence of large off-resonance effects. In fact, this limita
tion is exacerbated in INADEQUATE CR because of a more complex pulse sequen
ce. This paper describes the application to INADEQUATE CR of a general sche
me developed by Levitt to compensate an entire pulse sequence for pulse imp
erfections such as off-resonance effects. By incorporation of this scheme i
nto both INADEQUATE and INADEQUATE CR, it is shown that the doubled sensiti
vity of the latter is retained for camphor (covering about 200ppm) on a 500
MHz NMR spectrometer. Copyright (C) 2000 John Wiley & Sons, Ltd.