Effects of charge sharing in 3-D position sensitive CdZnTe gamma-ray spectrometers

Citation
Z. He et al., Effects of charge sharing in 3-D position sensitive CdZnTe gamma-ray spectrometers, NUCL INST A, 439(2-3), 2000, pp. 619-624
Citations number
6
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
439
Issue
2-3
Year of publication
2000
Pages
619 - 624
Database
ISI
SICI code
0168-9002(20000111)439:2-3<619:EOCSI3>2.0.ZU;2-M
Abstract
Two 3-D position-sensitive semiconductor gamma-ray spectrometers operable a t room temperature, have been fabricated using 1 cm(3) cubic CdZnTe crystal s. gamma-Ray spectra have been obtained from 11(x) x 11(y) x 20(z) voxels, each corresponding to similar to 0.7 x 0.7 x 0.5 mm in the detector volume. Individual spectra from each of the 2420 voxels were corrected for electro n trapping, differences in weighting potentials, and the gain variation of the readout circuitry. Energy resolutions of 1.70% (11.3 keV) FWHM and 1.84 % (12.2 keV) FWHM were obtained at 662 keV from the combined output of the entire volume for single-pixel events. Charge sharing between neighboring p ixels has been observed and the sizes of electron clouds are estimated from these experimental data. An understanding of electron cloud dimensions is important for the design of future detectors using pixellated electrodes. ( C) 2000 Elsevier Science B.V. All rights reserved.