Two 3-D position-sensitive semiconductor gamma-ray spectrometers operable a
t room temperature, have been fabricated using 1 cm(3) cubic CdZnTe crystal
s. gamma-Ray spectra have been obtained from 11(x) x 11(y) x 20(z) voxels,
each corresponding to similar to 0.7 x 0.7 x 0.5 mm in the detector volume.
Individual spectra from each of the 2420 voxels were corrected for electro
n trapping, differences in weighting potentials, and the gain variation of
the readout circuitry. Energy resolutions of 1.70% (11.3 keV) FWHM and 1.84
% (12.2 keV) FWHM were obtained at 662 keV from the combined output of the
entire volume for single-pixel events. Charge sharing between neighboring p
ixels has been observed and the sizes of electron clouds are estimated from
these experimental data. An understanding of electron cloud dimensions is
important for the design of future detectors using pixellated electrodes. (
C) 2000 Elsevier Science B.V. All rights reserved.