Objective: To compare examination time and visual field loss for ptosis fie
lds obtained with manual kinetic (Goldmann) perimetry and automated static
(Humphrey) perimetry.
Methods: Both eyes of 12 patients with bilateral aponeurogenic ptosis were
prospectively examined using Goldmann and Humphrey (ptosis protocol) perime
try with the eyelids ptotic and taped into a normal position.
Results: Bilateral examination time for Goldmann fields was 10 +/- 2 minute
s and for Humphrey fields was 50 +/- 10 minutes (P<.001, n = 12). Superior
fields at the 12:00 meridian were 46 degrees +/- 6 degrees taped, and 28 de
grees +/- 12 degrees untaped for Goldmann perimetry (P<.001), and 38 degree
s +/- 8 degrees taped, and 24 degrees +/- 12 degrees untaped for Humphrey p
erimetry P<.001). Goldmann field loss was 18 degrees +/- 9 degrees (taped m
inus untaped). Humphrey field loss was 14 degrees +/- 13 degrees (P<.04, n
= 24). Mean Goldmann radial fields were 56 degrees +/- 6 degrees taped and
39 degrees +/- 13 degrees untaped (P<.001). Goldmann superior hemifield are
as were 5167 +/- 964 degrees' taped and 2830 +/- 1466 degrees' untaped (P<.
001). Humphrey mean vertical superior hemifield was 37 degrees +/- 9 degree
s taped and 21 degrees +/- 11 degrees untaped (P<.001). Mean sensitivity of
Humphrey fields was 15 +/- 3 dB taped and 9 +/- 5 dB untaped (P<.001). Mea
n vertical center of gravity was 23 degrees +/- 3 degrees taped and 16 degr
ees +/- 5 degrees untaped (P<.001).
Conclusion: Goldmann manual kinetic and Humphrey automated static visual fi
eld testing are both effective in documenting ptosis associated visual fiel
d loss. Humphrey automated ptosis fields, as performed in this study, requi
re longer examination times than Goldmann manual fields and may be a less s
ensitive indicator of field loss.