Pj. Young et Mc. Blunt, Compliance characteristics of the Porter Soft Seal cuff improves seal against leakage of fluid in a pig trachea model, CRIT CARE, 3(5), 1999, pp. 123-126
Background: The Porter Soft Seal high-volume, low-pressure cuffed tracheal
tube was compared with the Mallinckrodt HiLo, Sheridan Preformed and Porter
Profile tracheal tubes for leakage of dye placed in the subglottic space o
f a pig's trachea which was used in a benchtop mechanical ventilation model
and in six isolated pig tracheas.
Results: There was no leakage, either in the ventilation model or in the is
olated tracheas in the Porter Soft Seal group, There was rapid leakage in t
he ventilation model and in all the isolated tracheas for the Mallinckrodt
HiLo, and five out of six isolated tracheas for the Sheridan Preformed and
the Porter Profile group.
Conclusions: This benchtop study suggests that the improved compliance char
acteristics of the Porter Soft Seal cuff are beneficial in preventing leaka
ge of fluid in these models.