Simultaneous determination of both the calibration constant in an electrochemical quartz crystal microbalance and the active surface area of a polycrystalline gold electrode
Ny. Gu et al., Simultaneous determination of both the calibration constant in an electrochemical quartz crystal microbalance and the active surface area of a polycrystalline gold electrode, ELECTROCH C, 2(1), 2000, pp. 48-50
A novel method is employed for the simultaneous determination of both the c
alibration constant of an electrochemical quartz crystal microbalance (EQCM
) and the active surface area of a polycrystalline gold electrode. A gold e
lectrode: is immersed into a 1 mM KI/1 M H2SO4 solution and on which forms
a neutral monolayer. The adsorbed iodine can then be completely oxidized in
to IO3-. The active surface area of a gold electrode can be obtained from t
he net electrolytic charge of the oxidation process, and the calibration co
nstant in the EQCM can be calculated from the corresponding frequency shift
. The result shows that this method is simple, convenient and valid. (C) 20
00 Elsevier Science S.A. All rights reserved.