Size and positioning reproducibility of an Ir-192 brachytherapy stepping source

Citation
A. Berndt et al., Size and positioning reproducibility of an Ir-192 brachytherapy stepping source, MED PHYS, 27(1), 2000, pp. 129-131
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MEDICAL PHYSICS
ISSN journal
00942405 → ACNP
Volume
27
Issue
1
Year of publication
2000
Pages
129 - 131
Database
ISI
SICI code
0094-2405(200001)27:1<129:SAPROA>2.0.ZU;2-Z
Abstract
In this paper we describe techniques for measuring the dimensions and posit ion reproducibility of an Ir-192 brachytherapy stepping source. Measurement s were carried out using a 0.25 X 10 X 152 mm(3) collimator placed in front of a detector of our own design. The brachytherapy source was translated p ast the collimator in 0.025 mm increments using a stepper motor. The source was found to be 3.58 mm long and 0.69 mm wide, which is in good agreement with the manufacturer's values of 3.5x0.6 mm(2). The source position was re producible to within 0.12 mm. (C) 2000 American Association of Physicists i n Medicine.