CHANNEL THICKNESS VARIATIONS COULD DEGRADE RESOLUTION IN ELECTROPHORESIS

Authors
Citation
G. Holzwarth, CHANNEL THICKNESS VARIATIONS COULD DEGRADE RESOLUTION IN ELECTROPHORESIS, Electrophoresis, 17(10), 1996, pp. 1587-1589
Citations number
11
Categorie Soggetti
Biochemical Research Methods
Journal title
ISSN journal
01730835
Volume
17
Issue
10
Year of publication
1996
Pages
1587 - 1589
Database
ISI
SICI code
0173-0835(1996)17:10<1587:CTVCDR>2.0.ZU;2-R
Abstract
Variations in the lateral dimensions of capillaries and ultrathin gels along their length could lead to a distortion of the sample zone beca use charged particles traveling near the wall must travel a longer dis tance under a smaller field than particles traveling along the centerl ine. The size of the effect is computed by solving Laplace's equation for the potential everywhere in a conducting slab with a discontinuity in the lateral dimensions and integrating the travel time along diffe rent streamlines.