The temperature calibration of a TA Instruments 3200-2920 DSC has been
performed on cooling using the isotropic --> nematic, isotropic --> c
holesteric and other liquid crystal --> liquid crystal transitions of
thermally stable, high purity liquid crystals. The thermal stability o
f these liquid crystals has been verified by measuring the temperature
of the mentioned transitions during cyclic heating and cooling experi
ments. Correspondence has been established between the real and indica
ted temperature during cooling for all combinations of heating and coo
ling rates of practical interest: correction values were determined to
the indicated temperature in order to obtain the real temperature on
cooling. These correction values were calculated as the average from t
he temperatures of four or five different liquid crystal transitions f
or each heating-cooling rate combination. The accuracy of the temperat
ure calibration on cooling is ca. 0.2 degrees C for heating and coolin
g rates up to 20 degrees C mid(-1).