J. Dominguez et al., Evaluation of a rapid immunochromatographic assay for the detection of Legionella antigen in urine samples, EUR J CL M, 18(12), 1999, pp. 896-898
Citations number
10
Categorie Soggetti
Microbiology
Journal title
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES
A new immunochromatographic membrane assay for detecting Legionella pneumop
hila serogroup 1 antigen in urine samples (Binax Now Legionella Urinary Ant
igen Test; Binax, USA) was evaluated. Its sensitivity, specificity and leve
l of agreement with the Binax enzyme immunoassay were compared using noncon
centrated and concentrated urine samples. The overall agreement between the
two tests was 98.1%: the specificity of both was 100%. The sensitivity of
the immunochromatographic assay was 55.5% in nonconcentrated urine and 97.2
% in concentrated urine in comparison with the enzyme immunoassay, using co
ncentrated urine as the reference test. This immunochromatographic assay sc
reens successfully for Legionella pneumophila serogroup 1 soluble antigen i
n concentrated urine samples.