Ep. Benis et Tjm. Zouros, Improving the energy resolution of a hemispherical spectrograph using a paracentric entry at a non-zero potential, NUCL INST A, 440(2), 2000, pp. 462-465
We demonstrate by ray tracing analysis that the focusing properties of a he
mispherical spectrograph depend critically on the placement of the electron
entrance position R-0 and the value of the potential V-0 at R-0. Improveme
nts over conventional spectrographs using R-0 = (R) over bar, the mean radi
us and V-0 = 0 are predicted for particular combinations of R-0 and V-0. A
spectrograph with (R) over bar = 101.6 mm and an entrance at R-0 = 82.55 mm
using a zoom lens and a two-dimensional position sensitive detector (2D-PS
D) with 40 mm multichannel plates and resistive anode encoder has been used
to confirm the predicted qualitative behaviour. These results suggest that
improvement in energy resolution may be attained without the use of fringi
ng field correctors. (C) 2000 Elsevier Science B.V. All rights reserved.