The dosimetric characteristics of gamma-ray, x-ray and electron irradiated
GAF-DM-1260 radiochromic film have been studied, and the dependence of radi
ation-induced film absorbance on irradiation temperature and reading temper
ature and the changes of the absorption spectrum peaks at different reading
temperatures and absorbed doses (3 x 10(1) to 5 x 10(4) Gy) are reported.
It is shown that the responses are independent of the gamma-ray dose rate i
n the dose rate range considered in the experiment. The film response chara
cteristics as a function of the absorbed dose to Co-60 gamma irradiation at
the spectrophotometric reading wavelengths of 400, 580, 600, 650 and 670 n
m and two absorption peaks are determined as well as the response character
istics to the electron beam at the reading wavelength of 400 nm. The result
s demonstrate that the film responses to gamma rays, x-rays and a 3.8 MeV e
lectron beam are equivalent, over the absorbed dose rate range of about 0.8
Gy.s(-1) to 5 x 10(8) Gy.s(-1). Some advice and points of view about the d
osimetric characteristics of the film and some problems in usage are provid
ed according to the experimental results. (C) 2000 Elsevier Science Ltd. Al
l rights reserved.