In order to circumvent the limitations of a quadrupole mass filter, which s
o far has been the most common mass analyzer for use with electrospray ioni
zation, a mass spectrometer coupling electrospray with time-of-flight mass
analysis has been constructed. The design of this spectrometer, including a
simple electrospray ion source, an atmosphere-vacuum interface without a r
f-only quadrupole/multipole that discriminates against low mass ions which
can be used for internal calibration, orthogonal extraction, and a reflectr
on, is presented together with its performance characteristics. A resolving
power of 7500 was obtained for the +5 ion of bovine insulin and the detect
ion limit for the same molecule was shown to be 1 f mol. Mass accuracy usin
g a conventional two-point calibration is 60 ppm for internal calibration o
n low mass ions and better than 100 ppm for external calibration on peptide
ions. An improved mass accuracy equal to 10 ppm was achieved using interna
l three-point calibration on low mass ions naturally present in the spectra
. The dynamic range of a single sample was found to be at least 400:1. (C)
2000 American Institute of Physics. [S0034-6748(00)02101-8].