LEAKAGE OF FLUID AROUND LOW-PRESSURE TRACHEAL TUBE CUFFS

Citation
M. Oikkonen et U. Aromaa, LEAKAGE OF FLUID AROUND LOW-PRESSURE TRACHEAL TUBE CUFFS, Anaesthesia, 52(6), 1997, pp. 567-569
Citations number
6
Categorie Soggetti
Anesthesiology
Journal title
ISSN journal
00032409
Volume
52
Issue
6
Year of publication
1997
Pages
567 - 569
Database
ISI
SICI code
0003-2409(1997)52:6<567:LOFALT>2.0.ZU;2-D
Abstract
The aim of the study was to evaluate leakage of liquid past the low-pr essure cuffs of tracheal tubes. Ten samples of each of nine different types oi tubes were tested in a PVC mock trachea, using intracuff pres sures of 20, 30, 40 and 50 cmH(2)O. In five types of tubes, 6-10 cuffs allowed profuse leakage (> 20 ml water in 5 min) even at the highest intracuff pressure, i.e. 50 cmH(2)O. In the most efficient tube, all t he cuffs were leak-proof (leakage < 5 ml in 5 min) at 40 cmH(2)O and i n the second best type the cuffs were leak-proof at 50 cmH(2)O. The le akage of fluid past the tracheal tube remains an unresolved problem wi th low-pressure cuffs.