Ion isolation in a continuous zero angle reflecting time-of-flight mass spectrometer

Authors
Citation
Cd. Hanson, Ion isolation in a continuous zero angle reflecting time-of-flight mass spectrometer, ANALYT CHEM, 72(3), 2000, pp. 448-453
Citations number
19
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICAL CHEMISTRY
ISSN journal
00032700 → ACNP
Volume
72
Issue
3
Year of publication
2000
Pages
448 - 453
Database
ISI
SICI code
0003-2700(20000201)72:3<448:IIIACZ>2.0.ZU;2-4
Abstract
A continuous zero angle reflecting time-of-flight mass spectrometer capable of tandem mass spectrometry measurements with high resolution and high sen sitivity has been developed. The instrument design features two pulsed-ion mirrors in a coaxial geometry. Ions can bet reflected back and forth with t he mirrors, which increases the net flight length and permits kinetic energ y focusing for enhanced resolution. The instrument also contains an electro static particle guide which increases ion transmission efficiency and can b e used in a bipolar pulsed mode to isolate ions of interest for structural study.