Miniaturised expanded-bed column with low dispersion suitable for fast flow-ELISA analyses

Citation
E. Palsson et al., Miniaturised expanded-bed column with low dispersion suitable for fast flow-ELISA analyses, BIOTECH LET, 22(3), 2000, pp. 245-250
Citations number
14
Categorie Soggetti
Biotecnology & Applied Microbiology",Microbiology
Journal title
BIOTECHNOLOGY LETTERS
ISSN journal
01415492 → ACNP
Volume
22
Issue
3
Year of publication
2000
Pages
245 - 250
Database
ISI
SICI code
0141-5492(200002)22:3<245:MECWLD>2.0.ZU;2-X
Abstract
Flow-ELISA measurements of the monoclonal antibody concentration in cultiva tion broth containing murine hybriboma cells were carried out using a small expanded-bed column (0.5x2.5 cm) charged with protein A. A new specialised pellicular agarose/stainless steel matrix designed for high flow rates wit h fast mass transport properties was used. Special care was taken to get an efficient flow distribution. The axial dispersion coefficient was very low (2x10(-6) m(2) s(-1) for latex particles at a linear velocity of 10 cm min (-1)). Breakthrough curves for polyclonal IgG on the protein A-derivatised support (at 2-11 cm min(-1)) further emphasised its advantageous properties . No significant change in dynamic capacity was found over the entire speed range.