Performance test of electron cyclotron resonance ion sources for the HyogoIon Beam Medical Center

Citation
K. Sawada et al., Performance test of electron cyclotron resonance ion sources for the HyogoIon Beam Medical Center, REV SCI INS, 71(2), 2000, pp. 987-989
Citations number
2
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
71
Issue
2
Year of publication
2000
Part
2
Pages
987 - 989
Database
ISI
SICI code
0034-6748(200002)71:2<987:PTOECR>2.0.ZU;2-W
Abstract
Two electron cyclotron resonance (ECR) ion sources were manufactured for th e accelerator facility at the Hyogo Ion Beam Medical Center. H-2(+), He2+, and C4+ were chosen as the accelerating ions because they have the highest charge to mass ratio among ion states which satisfy the required intensity and quality. The sources have the same structure as the 10 GHz ECR source a t the Heavy Ion Medical Accelerator in Chiba except for a few improvements in the magnetic structure. Their performance was investigated at the Sumito mo Heavy Industries factory before shipment. The maximum intensity was 1500 mu A for H-2(+), 1320 mu A for He2+, and 580 mu A for C4+ at the end of th e ion source beam transport line. These are several times higher than requi red. Sufficient performance was also observed in the flatness and long-term stability of the pulsed beams. These test results satisfy the requirements for medical use. (C) 2000 American Institute of Physics. [S0034-6748(00)53 802-7].