Highly-sensitive flow injection determination of hydrogen peroxide with a peroxidase-immobilized electrode and its application to clinical chemistry

Citation
K. Yamamoto et al., Highly-sensitive flow injection determination of hydrogen peroxide with a peroxidase-immobilized electrode and its application to clinical chemistry, ANALYT CHIM, 406(2), 2000, pp. 201-207
Citations number
29
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
406
Issue
2
Year of publication
2000
Pages
201 - 207
Database
ISI
SICI code
0003-2670(20000221)406:2<201:HFIDOH>2.0.ZU;2-Z
Abstract
A flow injection system for determination of hydrogen peroxide was construc ted using a peroxidase/ferrocene-embedded carbon paste (POD-Fc-CP) electrod e covered with a membrane. H2O2 was reduced bioelectrocatalytically at 100 mV (versus Ag/AgCl/sat. KCl). A Nafion(R)-coated cellulose acetate membrane prepared on the surface of water proved to be suitable for the covering me mbrane. The detection limit (signal :noise = 6) for H2O2 was 200 fmol (10 n M x 20 mu l) and the response linearity was retained at least up to 25 mu M . The relative standard deviation was 4.4% for 1.0 mu M H2O (n = 7). The fl ow injection system was successfully applied to the determination of glucos e and urate in serum by using the corresponding oxidase-immobilized pre-col umn. The detection limits were 1.5 pmol for glucose and 2 pmol for urate. ( C) 2000 Elsevier Science B.V. All rights reserved.