A clinical use of the Kyocera Gyro C1E3 centrifugal pump

Citation
M. Hata et al., A clinical use of the Kyocera Gyro C1E3 centrifugal pump, ARTIF ORGAN, 24(1), 2000, pp. 76-78
Citations number
8
Categorie Soggetti
Research/Laboratory Medicine & Medical Tecnology
Journal title
ARTIFICIAL ORGANS
ISSN journal
0160564X → ACNP
Volume
24
Issue
1
Year of publication
2000
Pages
76 - 78
Database
ISI
SICI code
0160-564X(200001)24:1<76:ACUOTK>2.0.ZU;2-0
Abstract
The Kyocera Gyro C1E3 centrifugal blood pump was clinically applied for a c ardiopulmonary bypass (CPB) of coronary artery bypass grafting (CABG). The patient was 72-year-old male with postinfarction unstable angina. The surge ry was carried out on November 20, 1998. The air inside the pump was easily and quickly re moved, and its controllability was excellent. The pump flow during operation was maintained 2.2 L/m(2). Total CPB time was 173 min. Pe rioperative parameters of hemolysis and cytotoxicity were not remarkably ch anged. Macroscopically and microscopically, there were no thrombi inside th e pump after usage. This is the first reported case of clinical use of the Kyocera Gyro C1E3 pump.