Complex and lengthy interventional radiological techniques have resulted in
a number of patients developing skin reactions in recent years. To safegua
rd against these side effects, we have investigated the degree to which ent
rance skin dose can be reduced by inserting 0.18 mm and 0.35 mm copper filt
ration in the incident beam. The potential reduction was measured on a 22 c
m water phantom for each of eight models of a fluoroscopy unit. Using the c
atheter laboratory fluoroscopy unit on which radiofrequency ablations are r
outinely performed, we assessed the relative effectiveness of adding filtra
tion and increasing the kV:mA ratio. Image quality was subjectively assesse
d for diagnostic and therapeutic acceptability in two groups of 10 patients
undergoing radiofrequency ablations, pacemaker insertions or electrophysio
logy studies. One of the groups was screened with 0.35 mm copper filtration
in place and the other group acted as the control. Maximum patient skin do
se proved difficult to measure directly because of the unpredictable dose p
attern. This pattern was studied in four patients using a film method in co
njunction with thermoluminescent dosemeters. Copper filtration 0.35 mm thic
k inserted in the beams of the eight fluoroscopy units produced a mean redu
ction in entrance dose to the phantom of 58% with a mean increase in tube l
oading of 29%. At 100 kV the increased loading on the X-ray tube was equiva
lent to increasing the anteroposterior separation of the patient by 2 cm. M
easurements on the catheter laboratory unit showed that the tube voltage wo
uld need to be raised above the normal diagnostic range to obtain an equiva
lent entrance dose reduction without the filter. The blackening of films un
der the patients showed complex patterns, but the estimated skin doses were
consistent with those predicted by the phantom experiments. All six cardio
logists considered there to be insignificant detriment to image quality in
the procedures investigated.