Primary measurement of ultrasonic power and dissemination of ultrasonic power reference values by means of standard transducers

Authors
Citation
K. Beissner, Primary measurement of ultrasonic power and dissemination of ultrasonic power reference values by means of standard transducers, METROLOGIA, 36(4), 1999, pp. 313-320
Citations number
17
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
METROLOGIA
ISSN journal
00261394 → ACNP
Volume
36
Issue
4
Year of publication
1999
Pages
313 - 320
Database
ISI
SICI code
0026-1394(1999)36:4<313:PMOUPA>2.0.ZU;2-Q
Abstract
The time-average ultrasonic power emitted by an ultrasonic transducer is on e of the key values characterizing the strength of ultrasonic fields as a w hole. For medical applications in particular, measurements are usually carr ied out in water and in the frequency range 1 MHz to 10 MHz. The use of a r adiation force balance is recommended in International Electrotechnical Com mission Standard TEC 61161. This measurement method and its practical reali zation at the Physikalisch-Technische Bundesanstalt are described. Uncertai nties involved in the primary measurement and in the dissemination process are discussed.