Summary of a European comparison of ultrasonic power measurements

Authors
Citation
K. Beissner, Summary of a European comparison of ultrasonic power measurements, METROLOGIA, 36(4), 1999, pp. 327-330
Citations number
6
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
METROLOGIA
ISSN journal
00261394 → ACNP
Volume
36
Issue
4
Year of publication
1999
Pages
327 - 330
Database
ISI
SICI code
0026-1394(1999)36:4<327:SOAECO>2.0.ZU;2-6
Abstract
A summary is given of the results of a European comparison of ultrasonic po wer measurements. Three national metrology institutes participated. All mea surements were carried out using the radiation force balance method accordi ng to International Electrotechnical Commission Standard IEC 61161, with mi nor differences in certain experimental details. Five standard transducers were circulated, with nominal resonance frequencies between 1 MHz and 10 MH z. The measurements covered the power range from IO mW to 200 mW, which rou ghly corresponds to the output of medical diagnostic equipment. The results obtained by the participants agreed within the estimated uncertainties.