A summary is given of the results of a European comparison of ultrasonic po
wer measurements. Three national metrology institutes participated. All mea
surements were carried out using the radiation force balance method accordi
ng to International Electrotechnical Commission Standard IEC 61161, with mi
nor differences in certain experimental details. Five standard transducers
were circulated, with nominal resonance frequencies between 1 MHz and 10 MH
z. The measurements covered the power range from IO mW to 200 mW, which rou
ghly corresponds to the output of medical diagnostic equipment. The results
obtained by the participants agreed within the estimated uncertainties.