Spin filters and supermirrors: a comparison study of two methods of high-precision neutron polarisation analysis

Citation
O. Zimmer et al., Spin filters and supermirrors: a comparison study of two methods of high-precision neutron polarisation analysis, NUCL INST A, 440(3), 2000, pp. 764-771
Citations number
19
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
440
Issue
3
Year of publication
2000
Pages
764 - 771
Database
ISI
SICI code
0168-9002(20000211)440:3<764:SFASAC>2.0.ZU;2-B
Abstract
The precise knowledge of the neutron polarisation is needed in tests of the electroweak Standard Model using angular correlations in polarised neutron beta decay. We performed an experimental comparison study of two different methods of polarisation analysis of a cold neutron beam which are based on spin-dependent reflection, respectively, transmission. The compare devices are a supermirror analyser used in prior neutron decay studies and an opaq ue transmission spin filter of either polarised He-3 or polarised protons. The results of the neutron polarisation measured with the supermirror analy ser and with the spin filter coincided in three different experiments withi n 0.1-0.2%. (C) 2000 Elsevier Science B.V. All rights reserved.