Flow-injection biamperometry of phenol

Citation
A. Kojlo et al., Flow-injection biamperometry of phenol, ANAL LETTER, 33(2), 2000, pp. 237-248
Citations number
18
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICAL LETTERS
ISSN journal
00032719 → ACNP
Volume
33
Issue
2
Year of publication
2000
Pages
237 - 248
Database
ISI
SICI code
0003-2719(2000)33:2<237:FBOP>2.0.ZU;2-Z
Abstract
A simple and rapid method determination of phenol is described, using a bro mination reaction and biamperometric detection. The method is based on meas urements of the intensity of current passing through two identical inert el ectrodes giving the concentration of phenol in samples. The sample is injec ted into a stream of carrier and merged with a stream of acidic bromine sol ution which brominates phenol in the reaction coil. A linear relationship b etween peak height and phenol concentration is observed in the range of 0.0 5 to 8 mu g ml(-1) of phenol. The relative standard deviation was 2.8% for analysis of 5 mu gml(-1) phenol and the sampling rate was 70 h(-1).