Determination of trace moisture in gases by diode-laser multi-pass absorption spectroscopy

Citation
Sq. Wu et al., Determination of trace moisture in gases by diode-laser multi-pass absorption spectroscopy, BUNSEKI KAG, 49(2), 2000, pp. 99-104
Citations number
14
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
BUNSEKI KAGAKU
ISSN journal
05251931 → ACNP
Volume
49
Issue
2
Year of publication
2000
Pages
99 - 104
Database
ISI
SICI code
0525-1931(200002)49:2<99:DOTMIG>2.0.ZU;2-J
Abstract
The amount of trace moisture is precisely determined in N-2 and HCl gases b y means of near-infrared diode-laser absorption spectroscopy. As an optical source, we use a near-infrared InGaAsP distributed-feedback diode laser wh ich operates at room temperature. A multi-pass sample cell designed for cor rosive gases is employed to enhance molecular absorption. For the purpose o f reducing the etalon noise that is usually encountered in multi-pass absor ption spectroscopy, we adjust the base path length in accordance with the b andwidth of a molecular spectral line so that the etalon frequency does not critically overlap with that of the molecular spectrum.. The remaining noi se is further reduced by applying multi-stage smoothing. Trace moisture con tent as small as 2.3 ppb is detected with a 20-m absorption length.