Investigations of artifact peaks in sensitive high-performance liquid chromatography methods

Citation
Gr. Strasser et I. Varadi, Investigations of artifact peaks in sensitive high-performance liquid chromatography methods, J CHROMAT A, 869(1-2), 2000, pp. 85-90
Citations number
5
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
Volume
869
Issue
1-2
Year of publication
2000
Pages
85 - 90
Database
ISI
SICI code
Abstract
Chromatograms of sensitive HPLC methods often show disturbing peaks, which mostly can be traced back easily. But sometimes such peaks are difficult to reproduce and therefore hard to trace. This paper describes two cases wher e artifact peaks arose due to contamination by the septum and by the sampli ng equipment, respectively, leading to misinterpretation of impurities and erroneous quantification. Furthermore, troubleshooting and ways to overcome these deficiencies are outlined. (C) 2000 Elsevier Science B.V. All rights reserved.