Repeatability and reproducibility of retention data and band profiles on reversed-phase liquid chromatography columns - IV. Results obtained with tuna C-18 (2) columns
M. Kele et G. Guiochon, Repeatability and reproducibility of retention data and band profiles on reversed-phase liquid chromatography columns - IV. Results obtained with tuna C-18 (2) columns, J CHROMAT A, 869(1-2), 2000, pp. 181-209
The reproducibility of retention data and band profile characteristics was
investigated for a series of columns packed with Luna C-18 (2), a silica-ba
sed, reversed-phase adsorbent. High precision data were obtained and statis
tically compared among five columns from the same batch (column-to-column r
eproducibility) and nine columns from as many different batches (batch-to-b
atch reproducibility). These data were acquired under five different sets o
f chromatographic conditions, for a group of 30 neutral, acidic, and basic
compounds selected as probes following an experimental protocol previously
described. This work is part of a study on the precision of chromatographic
analyses. Its purpose is to illustrate the contribution of the stationary
phase reproducibility to this precision. (C) 2000 Elsevier Science B.V. All
rights reserved.