Migration time correction for the analysis of derivatized amino acids and oligosaccharides by micellar capillary electrochromatography

Citation
Xf. Li et al., Migration time correction for the analysis of derivatized amino acids and oligosaccharides by micellar capillary electrochromatography, J CHROMAT A, 869(1-2), 2000, pp. 375-384
Citations number
16
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
Volume
869
Issue
1-2
Year of publication
2000
Pages
375 - 384
Database
ISI
SICI code
Abstract
Migration-time reproducibility is essential in the use of capillary electro phoresis to identify components in mixtures. Two methods based on the migra tion time of either one or two reference markers are proposed for improving migration time reproducibility. These methods were evaluated to determine the migration time reproducibility for phenylthiohydantoin-amino acids, flu orescein thiohydantoin-amino acids, and tetramethylrhodamine labeled oligos accharides. In the best case, the relative standard deviation of the migrat ion time was reduced from >3% without correction to <0.04% with the two-mar ker correction. (C) 2000 Elsevier Science B.V. All rights reserved.