S. Krishnaswami et al., A sensitive LC-MS/MS method for the quantification of fluticasone propionate in human plasma, J PHARM B, 22(1), 2000, pp. 123-129
A sensitive and selective LC-(APCI) MS/MS method capable of quantifying flu
ticasone propionate (FP) at levels dean to 10 pg ml(-1) in human plasma is
reported. The method was validated over a linear range from 10 to 1000 pg m
l(-1) using a previously published solid-phase extraction procedure with a
C-13(3)-labeled internal standard. The inter and intra:batch precision (coe
fficient of variation) and accuracy (% bias) of the quality controls sample
s (20, 25, 50, 100, 200, 500 and 1000 pg ml(-1)) were less than 15 and 11%,
respectively. The method is robust, rapid (analysis time of 2 min), select
ive and hence is ideally suited for pharmacokinetic investigations involvin
g inhalation of therapeutic doses of FP. (C) 2000 Elsevier Science B.V. All
rights reserved.