In the field of radioactivity measurement, there is a constant need for hig
hly specialized electronic modules such as ADCs, amplifiers, discriminators
, dead-time modules, etc. But sometimes it is almost impossible to find on
the market the modules having the performances corresponding to our needs.
The purpose of the module presented here. called MTR2 (Module de Temps-mort
Reconductible), is to process, in terms of pulse height discrimination and
dead-time corrections, the pulses delivered by the detectors used in count
ing systems. This dead-time, of the extendible type, is triggered by both t
he positive and negative parts of the incoming pulse and the dead-time corr
ections are made according to the live-time method. This module, which has
been developed and tested at LPRI, can be used alone in simple counting cha
nnels or in more complex systems such as coincidence systems. The philosoph
y governing the choice and the implementation of this type of dead-time as
well as the system used for the dead-time corrections is presented. The ele
ctronic scheme and the performances are also presented. This module is avai
lable in the NIM standard. (C) 2000 Elsevier Science Ltd. All rights reserv
ed.