Objective-To improve the biocompatibility of stents using a phosphorylcholi
ne coated stent as a form of biomimicry.
Interventions-Implantation of phosphorylcholine coated (n = 20) and non-coa
ted (n = 21) stents was performed in the coronary arteries of 25 pigs. The
animals were killed after five days (n = 6), four weeks (n = 7), and 12 wee
ks (n = 8), and the vessels harvested for histology, scanning electron micr
oscopy, and morphometry.
Main outcome measures-Stent performance was assessed by studying early endo
thelialisation, neointima formation, and vessel wall reaction to the synthe
tic coating.
Results-Stent thrombosis did not occur in either group. Morphometry showed
no significant differences between the two study groups at any time point.
At five days both the coated and non-coated stents were equally well endoth
elialised (91% v 92%, respectively). At four and 12 weeks there was no diff
erence in intimal thickness between the coated and non-coated stents. Up to
12 weeks postimplant the phosphorylcholine coating was still discernible i
n the stent strut voids, and did nut appear to elicit an adverse inflammato
ry response.
Conclusion-In this animal model the phosphorylcholine coating showed excell
ent blood and tissue compatibility, unlike a number of other polymers teste
d in a similar setting. Given that the coating was present up to 12 weeks p
ostimplant with no adverse tissue reaction, it may be a potential candidate
polymer for local drug delivery.