High-performance liquid chromatographic assay for minocycline in human plasma and parotid saliva

Citation
V. Orti et al., High-performance liquid chromatographic assay for minocycline in human plasma and parotid saliva, J CHROMAT B, 738(2), 2000, pp. 357-365
Citations number
29
Categorie Soggetti
Chemistry & Analysis
Journal title
JOURNAL OF CHROMATOGRAPHY B
ISSN journal
13872273 → ACNP
Volume
738
Issue
2
Year of publication
2000
Pages
357 - 365
Database
ISI
SICI code
1387-2273(20000211)738:2<357:HLCAFM>2.0.ZU;2-R
Abstract
A sensitive and specific high-performance liquid chromatographic (HPLC) met hod with UV detection was developed for the determination of minocycline in human plasma and parotid saliva samples. Samples were extracted using an O asis(TM) HLB cartridge and were injected into a C-8 Nucleosil column. The H PLC eluent contained acetonitrile-methanol-distilled water-0.1% trifluoroac etic acid (25:2:72.9:0.1, v/v). Demeclocycline was used as internal standar d. The assay showed linearity in the tested range of 0.1-25 mu g/ml. The li mit of quantitation was 100 ng/ml. Recovery from plasma or parotid saliva a veraged 95%. Precision expressed as %CV was in the range 0.2-17% (limit of quantitation). Accuracy ranged from 93 to 111%. In the two matrices studied at 20 and 4 degrees C, rapid degradation of the drug occurred. Frozen at - 30 degrees C, this drug was stable for at least 2 months, the percent recov ery averaged 90%. The method's ability to quantify minocycline with precisi on, accuracy and sensitivity makes it useful in pharmacokinetic studies. (C ) 2000 Elsevier Science B.V. All rights reserved.