Development and validation of a new high-performance liquid chromatographic estimation method of meloxicam in biological samples

Citation
T. Velpandian et al., Development and validation of a new high-performance liquid chromatographic estimation method of meloxicam in biological samples, J CHROMAT B, 738(2), 2000, pp. 431-436
Citations number
5
Categorie Soggetti
Chemistry & Analysis
Journal title
JOURNAL OF CHROMATOGRAPHY B
ISSN journal
13872273 → ACNP
Volume
738
Issue
2
Year of publication
2000
Pages
431 - 436
Database
ISI
SICI code
1387-2273(20000211)738:2<431:DAVOAN>2.0.ZU;2-G
Abstract
A simple, HPLC method was developed to estimate meloxicam (COX-2 inhibitor) using piroxicam as the internal standard. The mobile phase containing meth anol, acetonitrile and an aqueous solution of diammonium hydrogenorthophosp hate (50 mM) in the ratio of 4:1:5 was pumped at the rate 1 ml/min. Lichroc art RP-18 (125x4 mm) was used as an analytical column and the analytes were detected at 364 nm using a UV detector. Acidified plasma samples were extr acted with chloroform, evaporated to dryness, reconstituted in the mobile p hase and then a volume of 10 mu l of the prepared sample was injected in th e column. The retention time of meloxicam and piroxicam was found to be 2.7 and 1.9, respectively. This method showed an accuracy of 102.3% at 0.52 mu g/ml and was capable of detecting a minimum concentration of 0.029 mu g/ml meloxicam from biological samples. The analytical method was successfully utilized for estimating meloxicam in biological samples. (C) 2000 Elsevier Science B.V. All rights reserved.