A method developed at the Helsinki University of Technology (HUT) for calib
rating standard lamps of spectral irradiance in UV-A and UV-B regions is pr
esented. The method is based on a compact filter radiometer that is charact
erized absolutely to measure spectral irradiance. The filter radiometer can
be brought to laboratories where accurate calibrations are needed. The met
hod overcomes some of the instability problems encountered when using lamps
as transfer standards, which makes it also useful in intercomparison campa
igns. Test measurements are presented where two standard lamps issued by th
e National Institute of Standards and Technology (NIST), USA, are compared
with the HUT scale held by the radiometer. The results suggest that the agr
eement between the two scales (HUT and NIST) is approximately 1% - 2% in th
e wavelength region from 300 nm to 400 nm.