E. Pachart et al., Consistency tests on UV spectral irradiance measurements using modeling and a broadband instrument, J GEO RES-A, 105(D4), 2000, pp. 4851-4856
The UV spectroradiometer used at the University of Lille is presented, and
its characterization and calibration are described. In order to detect poss
ible shifts between calibrations, two different consistency tests are regul
arly performed on cloudless days. The erythemal dose rate obtained by integ
rating the weighted spectrum is compared to broadband measurements and mode
ling results. A Sun photometer provides the aerosol optical depth necessary
as an input to the radiative transfer model. Results of these tests are pr
esented for some cases; the general agreement is within 5%, which correspon
ds to the expected uncertainties.