An enhanced parallel port interface for controlling and monitoring experiments

Citation
Rc. Viscovini et al., An enhanced parallel port interface for controlling and monitoring experiments, MEAS SCI T, 11(2), 2000, pp. N23-N27
Citations number
6
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
MEASUREMENT SCIENCE & TECHNOLOGY
ISSN journal
09570233 → ACNP
Volume
11
Issue
2
Year of publication
2000
Pages
N23 - N27
Database
ISI
SICI code
0957-0233(200002)11:2<N23:AEPPIF>2.0.ZU;2-X
Abstract
In this paper we present a circuit for enhanced parallel port (EPP) interfa cing that is of low cost and great versatility for general use for data com munication in research laboratories. The communication is bi-directional an d its speed is higher than 500 Kbytes s(-1). This circuit has an address de coder, which permits individual communication with eight items of equipment of 8 and 16 bits, with an extra status bit. The addressing system also all ows insertion of up to 16 such interfacing circuits into the same computer EPP output, each of which can be independently reached.