Sa. Johansson et al., DOSIMETER GEL AND MR-IMAGING FOR VERIFICATION OF CALCULATED DOSE DISTRIBUTIONS IN CLINICAL RADIATION-THERAPY, Acta oncologica, 36(3), 1997, pp. 283-290
A dosimeter gel, based on an agarose gel infused with a ferrous sulpha
te solution and evaluated in a magnetic resonance scanner, was used fo
r complete Verification of calculated dose distributions. Two standard
treatment procedures, treatment of cancer in the urinary bladder and
treatment of breast cancer after modified radical mastectomy, were exa
mined using pixel-by-pixel and dose volume histogram comparison. The d
ose distributions calculated with the dose planning system was in very
good agreement with the measured ones. However, in the case of the mo
re complicated breast cancer treatment, some discrepancies were found,
mainly at the beam abutment region. This may be explained by held dis
placements errors and by a small limitation of the dose planning utili
sing small electron beams in this region. The dosimeter gel system hav
e proven to be a useful tool for dosimetry in clinical radiation thera
py applications.