Artifact size on three-dimensional (3D) magnetic resonance (MR) angiograms
and safety of various vascular clips (15 titanium and three absorbable poly
dioxanone clips) were assessed. All evaluated clips were completely safe. B
iodegradable clips rendered no artifacts; titanium clips were associated wi
th susceptibility effects. Artifact size was dependent on clip size, clip o
rientation, echo time, and degree of k-space coverage. In the presence of t
itanium vascular clips, fast 3D MR angiography should be performed with the
shortest echo time and full k-space coverage.