Mr. Recto et al., A technique to prevent newly implanted stent displacement during subsequent catheter and sheath manipulation, CATHET C IN, 49(3), 2000, pp. 297-300
We describe a novel technique to prevent the displacement or migration of a
newly implanted stent as a consequence of any subsequent catheter and shea
th manipulation during the same catheterization procedure. The technique in
volves reinflation of the dilation balloon within the stent immediately aft
er implant followed by advancing the long delivery sheath carefully over th
e balloon as the balloon is slowly deflated within the stent. The technique
was used successfully in 78 stents in 30 patients without stent dislodgmen
t or migration. (C) 2000 Wiley-Liss, Inc.