Method for the automated measurement of fine particle nitrate in the atmosphere

Citation
Mr. Stolzenburg et Sv. Hering, Method for the automated measurement of fine particle nitrate in the atmosphere, ENV SCI TEC, 34(5), 2000, pp. 907-914
Citations number
30
Categorie Soggetti
Environment/Ecology,"Environmental Engineering & Energy
Journal title
ENVIRONMENTAL SCIENCE & TECHNOLOGY
ISSN journal
0013936X → ACNP
Volume
34
Issue
5
Year of publication
2000
Pages
907 - 914
Database
ISI
SICI code
0013-936X(20000301)34:5<907:MFTAMO>2.0.ZU;2-2
Abstract
An integrated collection and vaporization cell has been developed to provid e automated, 10-min resolution monitoring of fine particle nitrate in the a tmosphere. Particles are collected by a humidified impaction process and an alyzed in place by flash vaporization and chemiluminescent detection of the evolved nitrogen oxides. Particle collection efficiency was measured betwe en 95% and 100% for particles above 0.1 mu m. Evaporative losses for ammoni um nitrate were 2 +/- 4%. Average nitrate values from the automated system were compared to traditional denuder filter measurements in three cities, w ith correlation coefficients of greater than 0.97 and regression slopes ran ging from 0.96 to 1.06. The detection limit is governed by the field blank, which for the Los Angeles area was 0.4 mu g/m(3). The system operated unat tended for days at a time, with data recovery of 97%.