This paper presents a practical microassembly process for three-dimensional
(3-D) microelectrode arrays for recording and stimulation in the central n
ervous system (CNS), Orthogonal lead transfers between the micromachined tw
o-dimensional probes and a cortical surface platform are formed by attachin
g gold beams on the probes to pads on the platform using wire-free ultrason
ic bonding. The low profile (150 mu m) outrigger design of the probes allow
s the bonding of fully assembled high-density arrays, Micromachined assembl
y tools allow the formation of a full 3-D probe array within 30 min. Arrays
having up to 8 x 16 shanks on 200-mu m centers have been realized and used
to record cortical single units successfully. Active 3-D probe arrays cont
aining on-chip CMOS signal processing circuitry have also been created usin
g the microassembly approach. In addition, a dynamic insertion technique ha
s been explored to allow the implantation of high-density probe arrays into
feline cortex at high-speed and with minimal traumatic injury.