Determination of quinolones in animal tissues and eggs by high-performanceliquid chromatography with photodiode-array detection

Citation
Pg. Gigosos et al., Determination of quinolones in animal tissues and eggs by high-performanceliquid chromatography with photodiode-array detection, J CHROMAT A, 871(1-2), 2000, pp. 31-36
Citations number
23
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
Volume
871
Issue
1-2
Year of publication
2000
Pages
31 - 36
Database
ISI
SICI code
Abstract
A rapid, specific reversed-phase HPLC method is described, with solid-phase extraction, for assaying five quinolones (ciprofloxacin, difloxacin, enrof loxacin, norfloxacin and marbofloxacin) with confirmative diode-array detec tion in samples of bovine kidney, muscle and eggs. The least efficient extr action was marbofloxacin from kidney tissue (64%). The lower detection limi t for each quinolone was: enrofloxacin and ciprofloxacin, 1 ng; norfloxacin and difloxacin, 2 ng; marbofloxacin, 4 ng injected. The intra-day relative standard deviations were lower than 7.9% and lower than 8.6% for inter-day assays. These results indicate that the developed method had an acceptable precision. (C) 2000 Elsevier Science B.V. All rights reserved.