Determination of sulphonated dyes in water by ion-interaction high-performance liquid chromatography

Citation
M. Perez-urquiza et al., Determination of sulphonated dyes in water by ion-interaction high-performance liquid chromatography, J CHROMAT A, 871(1-2), 2000, pp. 227-234
Citations number
16
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
Volume
871
Issue
1-2
Year of publication
2000
Pages
227 - 234
Database
ISI
SICI code
Abstract
An ion-interaction high-performance liquid chromatography method for quick separation and determination of the sulphonated dye Acid Yellow 1, and the sulphonated azo dyes Acid Orange 7, Acid Orange 12, Acid Orange 52, Acid Re d 2, Acid Red 26, Acid Red 27 and Acid Red ss has been developed. An RP-ODS stationary phase is used, and the mobile phase contains an acetonitrile-ph osphate buffer (27:73, v/v) mixture at pH 6.7, containing 2.4 mM butylamine as ion-interaction reagent. Good separations were obtained using isocratic elution and spectrophotometric detection at 460 nm. The detection limits f or the eight dyes ranged from 7 to 28 mu g/l for an injection volume of 100 mu l. Spiked tap water samples (100 ml), containing different concentratio n levels (0.3-1.2 mu g/l) of the dyes were analyzed after acidification (pH 3) and preconcentration in disposable solid-phase extraction C-18 cartridg es. (C) 2000 Elsevier Science B.V. All rights reserved.