Programmable temperature vaporization interface for on-line trace-level enrichment - GC-MS of micropollutants in surface water

Citation
S. De Koning et al., Programmable temperature vaporization interface for on-line trace-level enrichment - GC-MS of micropollutants in surface water, J MICROCOL, 12(3), 2000, pp. 153-159
Citations number
12
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF MICROCOLUMN SEPARATIONS
ISSN journal
10407685 → ACNP
Volume
12
Issue
3
Year of publication
2000
Pages
153 - 159
Database
ISI
SICI code
1040-7685(2000)12:3<153:PTVIFO>2.0.ZU;2-O
Abstract
In the present study, a fully automated solid-phase extraction (SPE) system , a programmed temperature vaporization (PTV) interface, and gas chromatogr aphy (GC) with mass spectrometric (MS) detection are combined for the trace -level analysis of micropollutants in surface water. The PTV interface is s elected because of its reliability regarding large volume injection and its robustness when analyzing real-life samples. The mass spectrometer was use d both in the full-scan and the selected ion monitoring modes and gave dete ction limits of about 15-40 ng/L and 0.7-5 ng/L (7.5-mL samples), respectiv ely, for a series of NP-pesticides. (C) 2000 John Wiley & Sons, Inc.