A chemical analyzer will be reported based upon coupling liquid chromatogra
phy (LC) with a water mobile phase via a novel interface to high-speed capi
llary gas chromatography (GC). The interface preserves the temporal informa
tion of the LC separation, thus a comprehensive two-dimensional LC X GC sep
aration is obtained. The LC separation is in the 5-10 min range, while the
GC separations are in the 1-2 second range. For a 5 min LC separation, 150
GC separations are obtained if the GC separations are every 2 s, thus a tru
ly comprehensive LC X GC data density is obtained. The 100% water mobile ph
ase simplifies instrumental requirements, thus enhancing separation and det
ection strategies, thus broadening the scope of LC to on-line process analy
sis and field monitoring applications by improving safety and minimizing wa
ste. Volatile organic compounds are analyzed by LC X GC with enhanced chemi
cal selectivity. Thus, this comprehensive LC X GC instrument is a useful al
ternative to current methods that employ headspace analysis. Separations of
compounds such as benzene, toluene, ethyl benzene, o-xylene (BTEX), as wel
l as chloroform and methylene chloride are presented. Injected sample conce
ntration limits of detection are in the 40-200 parts-per-billion range. (C)
2000 John Wiley & Sons, Inc.