Simplex optimisation of a flow injection analysis (FIA) system to determine rare earth elements (REE) with Arsenazo III

Citation
Jm. Sanchez et al., Simplex optimisation of a flow injection analysis (FIA) system to determine rare earth elements (REE) with Arsenazo III, ANAL LETTER, 33(3), 2000, pp. 553-569
Citations number
28
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICAL LETTERS
ISSN journal
00032719 → ACNP
Volume
33
Issue
3
Year of publication
2000
Pages
553 - 569
Database
ISI
SICI code
0003-2719(2000)33:3<553:SOOAFI>2.0.ZU;2-R
Abstract
A Flow Injection Method (FIA) for the determination of Rare Earth Elements (REE) in mixed solutions is described and the parameters involved are optim ised mathematically by a modified simplex method (MSM). The performance of the system permits the determination of REE from solutions with pH's rangin g from 0 to 6. A linear calibration range of up to ca. 30 mg L-1 for-an inj ection volume of 70 mu L and a concentration of Arsenazo III equal to 1.22 mM is achieved representing a tenfold increase over previously reported val ues. The detection limit of the method is 0.70 mg L-1 for all the elements studied except Yb which is less sensitive (1.27 mg L-1). In the case of Yb, an increase in the linear range of up to ca 40 mg L-1 is achieved. The FIA method can be applied in the analysis of mixtures containing different pai rs of REE elements with a relative error of less than 5%.