Jm. Sanchez et al., Simplex optimisation of a flow injection analysis (FIA) system to determine rare earth elements (REE) with Arsenazo III, ANAL LETTER, 33(3), 2000, pp. 553-569
A Flow Injection Method (FIA) for the determination of Rare Earth Elements
(REE) in mixed solutions is described and the parameters involved are optim
ised mathematically by a modified simplex method (MSM). The performance of
the system permits the determination of REE from solutions with pH's rangin
g from 0 to 6. A linear calibration range of up to ca. 30 mg L-1 for-an inj
ection volume of 70 mu L and a concentration of Arsenazo III equal to 1.22
mM is achieved representing a tenfold increase over previously reported val
ues. The detection limit of the method is 0.70 mg L-1 for all the elements
studied except Yb which is less sensitive (1.27 mg L-1). In the case of Yb,
an increase in the linear range of up to ca 40 mg L-1 is achieved. The FIA
method can be applied in the analysis of mixtures containing different pai
rs of REE elements with a relative error of less than 5%.