A. Ramirez et al., An alternative to the reference method for testing the radiochemical purity of Tc-99(m)-tetrofosmin, NUCL MED C, 21(2), 2000, pp. 199-203
The recommended method for controlling the radiochemical purity (RCP) of Tc
-99(m)-labelled tetrofosmin is thin-layer chromatography (TLC) with silica
gel as the stationary phase and acetone:dichloromethane (35:65, v/v) as the
mobile phase. However, this test is time-consuming and uses a toxic substa
nce. We propose an alternative method to control RCP, based on a solid-phas
e extraction method (SPE) using a Sep Pak(R) C18 cartridge. We used both me
thods to analyse 35 samples and to determine the percentages of colloid, ra
diopharmaceutical and free Tc-99(m)-pertechnerate. Normal labelling process
es were modified to obtain a wide range of values. The range of RCP values
obtained was 63.4-96.5% (median = 93.8%) by the standard TLC method and 70.
1-96.4% (median = 94.1%) by the SPE method. A significant difference was ob
served only for the percentage of free Tc-99(m)-pertechnetate (P < 0.05). T
he agreement between the two methods, using 90% RCP as the limit, gave a ka
ppa index = +1 (P < 0.001), indicating total agreement between them. The pr
oposed method is useful to control RCP because it yields results that are i
n good agreement with those of the standard method, and because it is safer
and less time-consuming. ((C) 2000 Lippincott Williams & Wilkins).